A microchannel plate (MCP) is used to detect single particles (electrons, ions and neutrons) and photons (ultraviolet radiation and X-rays). It is closely related to an electron multiplier, as both intensify single particles or photons by the multiplication of electrons via secondary emission. Because a microchannel plate detector has many separate channels, it can provide spatial resolution.
Basic design A microchannel plate is a slab made from resistive material (most often glass) 0.5 to 2 mm thick with a regular array of tiny tubes (microchannels) leading from one face to the other. The microchannels are typically 5-20 micrometers in diameter, parallel to each other and enter the plate at a small angle to the surface (8-13° from normal). Plates are often round disks, but can be cut to any shape from sizes 10 mm up to 200 mm. They may also be curved.
Operating mode At non-relativistic energies, single particles generally produce effects too small to enable their direct detection. The microchannel plate functions as a particle amplifier, turning a single impinging particle into a cloud of electrons. By applying a strong electric field across the MCP, each individual microchannel becomes a continuous-dynode electron multiplier. A particle or photon that enters one of the channels through a small orifice is guaranteed to hit the wall of the channel, due to the channel being at an angle to the plate. The impact starts a cascade of electrons that propagates through the channel, amplifying the original signal by several orders of magnitude, depending on the electric field strength and the geometry of the microchannel plate. After the cascade, the microchannel takes time to recover (or recharge) before it can detect another signal. The electrons exit the channels on the opposite side of the plate, where they are collected on an anode. Some anodes are designed to allow spatially resolved ion collection, producing an image of the particles or photons incident on the plate. Although in many cases the collecting anode functions as the detecting element, the MCP itself can also be used as a detector. The discharging and recharging of the plate produced by the electron cascade, can be decoupled from the high voltage applied to the plate and measured, to directly produce a signal corresponding to a single particle or photon. The gain of an MCP is very noisy, meaning that two identical particles detected in succession will often produce wildly different signal magnitudes. The temporal jitter resulting from the peak height variation can be removed by using a constant fraction discriminator. Thusly employed, MCPs are capable of measuring particle arrival times with high resolution, making them ideal detectors for mass spectrometers.
Chevron MCP
Most modern MCP detectors consist of two microchannel plates with angled channels, rotated 180° from each other - producing a shallow chevron (v-like) shape. In a chevron MCP, the electrons that exit the first plate start the cascade in the next plate. The angle between the channels reduces ion feedback in the device, as well as producing significantly more gain at a given voltage, compared to a straight channel MCP. The two MCPs can either be pressed together to preserve spatial resolution, or have a small gap between them to spread the charge across multiple channels, which further increases the gain.
Z stack MCP This is an assembly of three microchannel plates with channels aligned in a Z shape. Single MCPs can have gain up to 10,000 (40 dB) but this system can provide gain more than 10 million (70 dB).
The detector
An external voltage divider is used to apply 100 volts to the acceleration optics (for electron detection), each MCP, the gap between the MCPs, the backside of the last MCP, and the collector (anode). The last voltage dictates the time of flight of the electrons and in this way, the pulse-width. The anode is a 0.4 mm thick plate with an edge of 0.2 mm radius to avoid high field strengths. It is just large enough to cover the active area of the MCP, because the backside of the last MCP, and the anode, together act as a capacitor with 2 mm separation - and large capacitance slows down the signal. The positive charge in the MCP influences positive charge in the backside metalization. A hollow torus conducts this around the edge of the anode plate. A torus is the optimum compromise between low capacitance and short path and for similar reasons, usually no dielectric (Markor) is placed into this region. After a 90° turn of the torus it is possible to attach a large coaxial waveguide. A taper permits minimizing the radius so that an SMA connector can be used. To save space and make the impedance match less critical, the taper is often reduced to a small 45° cone on the backside of the anode plate. The typical 500 volts between the backside of the last MCP and the anode cannot be fed directly into the preamplifier; the inner or the outer conductor needs a DC block, that is, a capacitor. Often it is chosen to only have 10-fold capacitance compared to the MCP-anode capacitance and is implemented as a plate capacitor. Rounded, electro-polished metal plates and the ultra high vacuum allow very high field strengths and high capacitance without a dielectric. The bias for the center conductor is applied via resistors hanging through the waveguide (see bias tee). If the DC block is used in the outer conductor, it is aligned in parallel with the larger capacitor in the power supply. Assuming good screening, the only noise is due to current noise from the linear power regulator. Because the current is low in this application and space for large capacitors is available, and because the DC-block capacitor is fast, it is possible to have very low voltage noise, so that even weak MCP signals can be detected. Sometimes the preamplifier is on a potential (off ground) and gets its power through a low-power isolation transformer and outputs its signal optically.