ArticleslgStudy

chemistry

Photoelectron photoion coincidence spectroscopy

Photoelectron photoion coincidence spectroscopy is a chemistry topic covered in the lgStudy science library. This page brings together a partial reference excerpt, illustrations, worked examples, real-world applications and a short study plan, so you can understand Photoelectron photoion coincidence spectroscopy rather than just read about it. In short: Photoelectron photoion coincidence spectroscopy (PEPICO) is a combination of photoionization mass spectrometry and photoelectron spectroscopy. It is largely based on the photoelectric effect.

Photoelectron photoion coincidence spectroscopy — main illustration
Photoelectron photoion coincidence spectroscopy — illustration

Key takeaways

  • Photoelectron photoion coincidence spectroscopy belongs to chemistry; place it in that map before memorising details.
  • Learn the definition first, then one example that makes the definition concrete.
  • Connect Photoelectron photoion coincidence spectroscopy to a quantity you can measure, compute or draw — that is where exam questions come from.
  • Reproduce the core statement of Photoelectron photoion coincidence spectroscopy from memory before moving on to harder problems.

Reference excerpt

Photoelectron photoion coincidence spectroscopy (PEPICO) is a combination of photoionization mass spectrometry and photoelectron spectroscopy. It is largely based on the photoelectric effect. Free molecules from a gas-phase sample are ionized by incident vacuum ultraviolet (VUV) radiation. In the ensuing photoionization, a cation and a photoelectron are formed for each sample molecule. The mass of the photoion is determined by time-of-flight mass spectrometry, whereas, in current setups, photoelectrons are typically detected by velocity map imaging. Electron times-of-flight are three orders of magnitude smaller than those of ions, which allows electron detection to be used as a time stamp for the ionization event, starting the clock for the ion time-of-flight analysis. In contrast with pulsed experiments, such as REMPI, in which the light pulse must act as the time stamp, this allows the use of continuous light sources, e.g. a discharge lamp or a synchrotron light source. No more than several ion–electron pairs are present simultaneously in the instrument, and the electron–ion pairs belonging to a single photoionization event can be identified and detected in delayed coincidence.

History

Brehm and von Puttkammer published the first PEPICO study on methane in 1967. In the early works, a fixed energy light source was used, and the electron detection was carried out using retarding grids or hemispherical analyzers: the mass spectra were recorded as a function of electron energy. Tunable vacuum ultraviolet light sources were used in later setups, in which fixed, mostly zero kinetic energy electrons were detected, and the mass spectra were recorded as a function of photon energy. Detecting zero kinetic energy or threshold electrons in threshold photoelectron photoion coincidence spectroscopy, TPEPICO, has two major advantages. Firstly, no kinetic energy electrons are produced in energy ranges with poor Franck–Condon factors in the photoelectron spectrum, but threshold electrons can still be emitted via other ionization mechanisms. Secondly, threshold electrons are stationary and can be detected with higher collection efficiencies, thereby increasing signal levels. Threshold electron detection was first based on line-of-sight, i.e. a small positive field was applied towards the electron detector, and kinetic energy electrons with perpendicular velocities are stopped by small apertures. The inherent compromise between resolution and collection efficiency was resolved by applying velocity map imaging conditions. Most recent setups offer meV or better (0.1 kJ mol−1) resolution both in terms of photon energy and electron kinetic energy. The 5–20 eV (500–2000 kJ mol−1, λ = 250–60 nm) energy range is of prime interest in valence photoionization. Widely tunable light sources are few and far between in this energy range. The only laboratory based one is the H2 discharge lamp, which delivers quasi-continuous radiation up to 14 eV. The few high resolution laser setups for this energy range are not easily tunable over several eV. Currently, VUV beamlines at third generation synchrotron light sources are the brightest and most tunable photon sources for valence ionization. The first high energy resolution PEPICO experiment at a synchrotron was the pulsed-field ionization setup at the Chemical Dynamics Beamline of the Advanced Light Source.

Overview

The primary application of TPEPICO is the production of internal energy selected ions to study their unimolecular dissociation dynamics as a function of internal energy. The electrons are extracted by a continuous electric field and are velocity map imaged depending on their initial kinetic energy. Ions are accelerated in the opposite direction and their mass is determined by time-of-flight mass spectrometry. The data analysis yields dissociation thresholds, which can be used to derive new thermochemistry for the sample. The electron imager side can also be used to record photoionization cross sections, photoelectron energy and angular distributions. With the help of circularly polarized light, photoelectron circular dichroism (PECD) can be studied. A thorough understanding of PECD effects could help explain the homochirality of life. Flash pyrolysis can also be used to produce free radicals or intermediates, which are then characterized to complement e.g. combustion studies. In such cases, the photoion mass analysis is used to confirm the identity of the radical produced. Photoelectron photoion coincidence spectroscopy can be used to shed light on reaction mechanisms, and can also be generalized to study double ionization in (photoelectron) photoion photoion coincidence ((PE)PIPICO), fluorescence using photoelectron photon coincidence (PEFCO), or photoelectron photoelectron coincidence (PEPECO). Times-of-flight of photoelectrons and photoions can be combined in a form of a map, which visualizes the dynamics of the dissociative ionization process. Ion–electron velocity vector correlation functions can be obtained in double imaging setups, in which the ion detector also delivers position information.

Energy selection

The relatively low intensity of the ionizing VUV radiation guarantees one-photon processes, in other words only one, fixed energy photon will be responsible for photoionization. The energy balance of photoionization comprises the internal energy and the adiabatic ionization energy of the neutral as well as the photon energy, the kinetic energy of the photoelectron and of the photoion. Because only threshold electrons are considered and the conservation of momentum holds, the last two terms vanish, and the internal energy of the photoion is known:

E int ion = E int neutral + h ν − I E ad {\displaystyle E_{\text{int}}^{\text{ion}}=E_{\text{int}}^{\text{neutral}}+h\nu -IE_{\text{ad}}}

… excerpt ends here. Continue reading the full article.

Illustrations

Photoelectron photoion coincidence spectroscopy: Velocity map imaging photoelectron photoion coincidence apparatus. Electrons with different kinetic energies are shown as well as ions with a room temperature kinetic energy distribution.
Velocity map imaging photoelectron photoion coincidence apparatus. Electrons with different kinetic energies are shown as well as ions with a room temperature kinetic energy distribution.
Photoelectron photoion coincidence spectroscopy: Potential energy diagram for dissociative photoionization. When only zero kinetic energy electrons are detected, the photon energy above the adiabatic ionization energy is converted into the internal energy of the photoion AB+
Potential energy diagram for dissociative photoionization. When only zero kinetic energy electrons are detected, the photon energy above the adiabatic ionization energy is converted into the internal energy of the photoion AB+

Worked examples

Example 1 — a first encounter with Photoelectron photoion coincidence spectroscopy

Start with the simplest possible case. Write down what Photoelectron photoion coincidence spectroscopy claims or describes in one sentence, then invent the smallest concrete situation in which that sentence is true. In chemistry, the smallest case is usually a single object, a single equation or a single measurement. Check that every symbol or term in your sentence has a meaning in that case.

Example 2 — changing one variable

Take the situation from Example 1 and change exactly one quantity: double it, halve it, or set it to zero. Predict what should happen to Photoelectron photoion coincidence spectroscopy before you calculate. Comparing your prediction with the result is the fastest way to find out whether you understand the idea or only the words.

Example 3 — an exam-style question

Typical questions about Photoelectron photoion coincidence spectroscopy ask you to (a) state it precisely, (b) apply it to given data, and (c) explain a limitation. Practise writing all three answers in under five minutes; the third part is what separates a full-mark answer from an average one.

Applications of Photoelectron photoion coincidence spectroscopy

In research
Photoelectron photoion coincidence spectroscopy appears in chemistry research whenever the underlying quantities have to be modelled precisely. Papers usually cite it as a starting assumption and then explore where it breaks down.
In technology and industry
Engineering practice reuses Photoelectron photoion coincidence spectroscopy in design rules, simulations and safety margins. Knowing the idea lets you read a specification sheet and understand why the numbers look the way they do.
In the classroom
Photoelectron photoion coincidence spectroscopy is common in secondary-school and first-year university syllabi. It links to neighbouring topics Chemical kinetics, Electron spectroscopy, Mass spectrometry, so understanding it makes those chapters shorter.
In everyday life
Look for Photoelectron photoion coincidence spectroscopy outside the textbook — in sport, cooking, traffic, electronics or the sky above you. An example you found yourself is remembered far longer than one you were given.
Ask Teacher Smith questions about this articleOpens your AI tutor with a question about “Photoelectron photoion coincidence spectroscopy” →

Affiliate

Preply — study more efficiently by working with a personal tutor. 50% off.

How to study Photoelectron photoion coincidence spectroscopy in 20 minutes

  1. Read the reference excerpt below once, without taking notes.
  2. Close the page and write down what Photoelectron photoion coincidence spectroscopy means in your own words.
  3. Compare your version with the excerpt and mark what you missed.
  4. Work through the three examples above with pen and paper.
  5. Explain Photoelectron photoion coincidence spectroscopy out loud to somebody else — or to Teacher Smith in the lgStudy chat.

Frequently asked questions

What is Photoelectron photoion coincidence spectroscopy in simple terms?

Photoelectron photoion coincidence spectroscopy (PEPICO) is a combination of photoionization mass spectrometry and photoelectron spectroscopy. It is largely based on the photoelectric effect.

Why does Photoelectron photoion coincidence spectroscopy matter?

Because it connects several chemistry ideas at once: it gives you a definition you can apply, a quantity you can calculate, and a way to check whether a result is plausible.

How should I study Photoelectron photoion coincidence spectroscopy?

Read the excerpt, restate it from memory, then work through the examples and applications listed on this page. The five-step study plan above takes about twenty minutes.

What does this page cover?

It gives you a compact reference excerpt plus original lgStudy explanations, examples, applications and study material on Photoelectron photoion coincidence spectroscopy.

Tags

  • Chemical kinetics
  • Electron spectroscopy
  • Mass spectrometry
  • Physical chemistry
  • Spectroscopy
  • Thermochemistry

Keep exploring