ArticleslgStudy

science

Scanning SQUID microscopy

Scanning SQUID microscopy is a science topic covered in the lgStudy science library. This page brings together a partial reference excerpt, illustrations, worked examples, real-world applications and a short study plan, so you can understand Scanning SQUID microscopy rather than just read about it. In short: In condensed matter physics, scanning SQUID microscopy is a technique where a superconducting quantum interference device (SQUID) is used to image surface magnetic field strength with micrometre-scale resolution. A tiny SQUID is mounted onto a tip which is then rastered near the surface of the sample to be measured.

Scanning SQUID microscopy — main illustration
Scanning SQUID microscopy — illustration

Key takeaways

  • Scanning SQUID microscopy belongs to science; place it in that map before memorising details.
  • Learn the definition first, then one example that makes the definition concrete.
  • Connect Scanning SQUID microscopy to a quantity you can measure, compute or draw — that is where exam questions come from.
  • Reproduce the core statement of Scanning SQUID microscopy from memory before moving on to harder problems.

Reference excerpt

In condensed matter physics, scanning SQUID microscopy is a technique where a superconducting quantum interference device (SQUID) is used to image surface magnetic field strength with micrometre-scale resolution. A tiny SQUID is mounted onto a tip which is then rastered near the surface of the sample to be measured. As the SQUID is the most sensitive detector of magnetic fields available and can be constructed at submicrometre widths via lithography, the scanning SQUID microscope allows magnetic fields to be measured with unparalleled resolution and sensitivity. The first scanning SQUID microscope was built in 1992 by Black et al. Since then the technique has been used to confirm unconventional superconductivity in several high-temperature superconductors including YBCO and BSCCO compounds.

Operating principles

The scanning SQUID microscope is based upon the thin-film DC SQUID. A DC SQUID consists of superconducting electrodes in a ring pattern connected by two weak-link Josephson junctions (see figure). Above the critical current of the Josephson junctions, the idealized difference in voltage between the electrodes is given by

V = R 2 I 2 − I 0 2 , = R 2 ( I 2 − ( 2 I c cos ⁡ ( π Φ Φ 0 ) ) 2 ) 1 2 , {\displaystyle {\begin{aligned}V&={\frac {R}{2}}{\sqrt {I^{2}-I_{0}^{2}}},\\&={\frac {R}{2}}\left(I^{2}-\left(2I_{c}\cos \left(\pi {\frac {\Phi }{\Phi _{0}}}\right)\right)^{2}\right)^{\frac {1}{2}},\end{aligned}}}

where R is the resistance between the electrodes, I is the current, I0 is the maximum supercurrent, Ic is the critical current of the Josephson junctions, Φ is the total magnetic flux through the ring, and Φ0 is the magnetic flux quantum. Hence, a DC SQUID can be used as a flux-to-voltage transducer. However, as noted by the figure, the voltage across the electrodes oscillates sinusoidally with respect to the amount of magnetic flux passing through the device. As a result, alone a SQUID can only be used to measure the change in magnetic field from some known value, unless the magnetic field or device size is very small such that Φ < Φ0. To use the DC SQUID to measure standard magnetic fields, one must either count the number of oscillations in the voltage as the field is changed, which is very difficult in practice, or use a separate DC bias magnetic field parallel to the device to maintain a constant voltage and consequently constant magnetic flux through the loop. The strength of the field being measured will then be equal to the strength of the bias magnetic field passing through the SQUID. Although it is possible to read the DC voltage between the two terminals of the SQUID directly, because noise tends to be a problem in DC measurements, an alternating current technique is used. In addition to the DC bias magnetic field, an AC magnetic field of constant amplitude, with field strength generating Φ << Φ0, is also emitted in the bias coil. This AC field produces an AC voltage with amplitude proportional to the DC component in the SQUID. The advantage of this technique is that the frequency of the voltage signal can be chosen to be far away from that of any potential noise sources. By using a lock-in amplifier the device can read only the frequency corresponding to the magnetic field, ignoring many other sources of noise.

… excerpt ends here. Continue reading the full article.

Illustrations

Scanning SQUID microscopy: Left: Schematic of a scanning SQUID microscope in a helium-4 refrigerator. Green holder for the SQUID probe is attached to a quartz tuning fork. Bottom part is a piezoelectric sample stage. Right: electron micrograph of a SQUID probe and a test image of Nb/Au strips recorded with it.[1]
Left: Schematic of a scanning SQUID microscope in a helium-4 refrigerator. Green holder for the SQUID probe is attached to a quartz tuning fork. Bottom part is a piezoelectric sample stage. Right: electron micrograph of a SQUID probe and a test image of Nb/Au strips recorded with it.[1]
Scanning SQUID microscopy: Diagram of a DC SQUID. The current 
  
    
      
        I
      
    
    {\displaystyle I}
  
 enters and splits into the two paths, each with currents 
  
    
      
        
          I
          
            a
          
        
      
    
    {\displaystyle I_{a}}
  
 and 
  
    
      
        
          I
          
            b
          
        
      
    
    {\displaystyle I_{b}}
  
. The thin barriers on each path are Josephson junctions, which together separate the two superconducting regions. 
  
    
      
        Φ
      
    
    {\displaystyle \Phi }
  
 represents the magnetic flux entering the inside of the DC SQUID loop.
Diagram of a DC SQUID. The current I {\displaystyle I} enters and splits into the two paths, each with currents I a {\displaystyle I_{a}} and I b {\displaystyle I_{b}} . The thin barriers on each path are Josephson junctions, which together separate the two superconducting regions. Φ {\displaystyle \Phi } represents the magnetic flux entering the inside of the DC SQUID loop.
Scanning SQUID microscopy: Scanning SQUID microscope
Scanning SQUID microscope
Scanning SQUID microscopy: Figure 1: Electrical schematic of a SQUID where Ib is the bias current, I0 is the critical current of the SQUID, Φ is the flux threading the SQUID and V is the voltage response to that flux.
Figure 1: Electrical schematic of a SQUID where Ib is the bias current, I0 is the critical current of the SQUID, Φ is the flux threading the SQUID and V is the voltage response to that flux.
Scanning SQUID microscopy: Figure 2 a) Plot of current vs. voltage for a SQUID. Upper and lower curves correspond to nΦ0 and (n+1/2)Φ0 respectively. Figure 2 b) Periodic voltage response due to flux through a SQUID. The periodicity is equal to one flux quantum, Φ0
Figure 2 a) Plot of current vs. voltage for a SQUID. Upper and lower curves correspond to nΦ0 and (n+1/2)Φ0 respectively. Figure 2 b) Periodic voltage response due to flux through a SQUID. The periodicity is equal to one flux quantum, Φ0

Worked examples

Example 1 — a first encounter with Scanning SQUID microscopy

Start with the simplest possible case. Write down what Scanning SQUID microscopy claims or describes in one sentence, then invent the smallest concrete situation in which that sentence is true. In science, the smallest case is usually a single object, a single equation or a single measurement. Check that every symbol or term in your sentence has a meaning in that case.

Example 2 — changing one variable

Take the situation from Example 1 and change exactly one quantity: double it, halve it, or set it to zero. Predict what should happen to Scanning SQUID microscopy before you calculate. Comparing your prediction with the result is the fastest way to find out whether you understand the idea or only the words.

Example 3 — an exam-style question

Typical questions about Scanning SQUID microscopy ask you to (a) state it precisely, (b) apply it to given data, and (c) explain a limitation. Practise writing all three answers in under five minutes; the third part is what separates a full-mark answer from an average one.

Applications of Scanning SQUID microscopy

In research
Scanning SQUID microscopy appears in science research whenever the underlying quantities have to be modelled precisely. Papers usually cite it as a starting assumption and then explore where it breaks down.
In technology and industry
Engineering practice reuses Scanning SQUID microscopy in design rules, simulations and safety margins. Knowing the idea lets you read a specification sheet and understand why the numbers look the way they do.
In the classroom
Scanning SQUID microscopy is common in secondary-school and first-year university syllabi. It links to neighbouring topics Josephson effect, Measuring instruments, Scanning probe microscopy, so understanding it makes those chapters shorter.
In everyday life
Look for Scanning SQUID microscopy outside the textbook — in sport, cooking, traffic, electronics or the sky above you. An example you found yourself is remembered far longer than one you were given.

Affiliate

Preply — study more efficiently by working with a personal tutor. 50% off.

How to study Scanning SQUID microscopy in 20 minutes

  1. Read the reference excerpt below once, without taking notes.
  2. Close the page and write down what Scanning SQUID microscopy means in your own words.
  3. Compare your version with the excerpt and mark what you missed.
  4. Work through the three examples above with pen and paper.
  5. Explain Scanning SQUID microscopy out loud to somebody else — or to Teacher Smith in the lgStudy chat.

Frequently asked questions

What is Scanning SQUID microscopy in simple terms?

In condensed matter physics, scanning SQUID microscopy is a technique where a superconducting quantum interference device (SQUID) is used to image surface magnetic field strength with micrometre-scale resolution. A tiny SQUID is mounted onto a tip which is then rastered near the surface of the samp…

Why does Scanning SQUID microscopy matter?

Because it connects several science ideas at once: it gives you a definition you can apply, a quantity you can calculate, and a way to check whether a result is plausible.

How should I study Scanning SQUID microscopy?

Read the excerpt, restate it from memory, then work through the examples and applications listed on this page. The five-step study plan above takes about twenty minutes.

What does this page cover?

It gives you a compact reference excerpt plus original lgStudy explanations, examples, applications and study material on Scanning SQUID microscopy.

Tags

  • Josephson effect
  • Measuring instruments
  • Scanning probe microscopy
  • Superconductivity

Keep exploring